Data sheet?
Number of input channels | 4-channel |
---|---|
Power supply | 24VDC ![]() |
Allowable voltage range | 90 to 110% of rated voltage |
Power consumption | Max. 5W(for max. load) |
Display method | None-parameter setting and monitoring is available at external devices(PC, PLC, etc.) |
Input type_TC | K(CA), J(IC), E(CR), T(CC), B(PR), R(PR), S(PR), N(NN), C(TT), G(TT), L(IC), U(CC), Platinel II |
Input type_RTD | DPt100?, JPt100?, DPt50?, Cu100?, Cu50?, Nikel 120? 3-wire type(permissible line resistance max. 5?? |
Input type_Analog | Voltage: 0-100mVDC ![]() ![]() ![]() ![]() Current: 0-20mA, 4-20mA |
Sampling cycle | 50ms(2-CH or 4-CH synchronous sampling) |
Measurement accuracy_Thermocouple | At room temperature(23?????: (PV ?0.3% or ?1?? higher one) ?1-digit Out of room temperature range: (PV ?0.5% or ?2?? higher one) ?1-digit |
Measurement accuracy_RTD | At room temperature(23?????: (PV ?0.3% or ?1?? higher one) ?1-digit Out of room temperature range: (PV ?0.5% or ?2?? higher one) ?1-digit |
Measurement accuracy_Analog | At room temperature(23?????: ?0.3% F.S. ?1-digit Out of room temperature range: ?0.5% F.S. ?1-digit |
Option input_CT input | 0.0-50.0A(primary current measurement range) ??T ratio=1/1000, Measured accuracy: ?5% F.S. ?1-digit |
Control method | ON/OFF control, P, PI, PD, PID control |
Control output_Relay | 250VAC~ 3A 1a |
Control output_SSR | Max. 12VDC ![]() |
Control output_Current | Selectable DC 4-20mA or DC 0-20mA(load resistance max. 500?? |
Option output_Alarm | 250VAC~ 3A 1a |
Communication_Comm. terminal | RS485 communication output(Modbus RTU) |
Communication_PC loader | TTL |
Hysteresis(adjustable sensitivity) | RTD/Thermocouples: 1 to 100????(0.1 to 100.0????, analog: 1 to 100 digit |
Proportional band | RTD/Thermocouples: 1 to 999????(0.1 to 999.9????, analog: 0.1 to 999.9 digit |
Integral time | 0 to 9999 sec |
Derivative time | 0 to 9999 sec |
Control period | Relay output: 0.1 to 120.0 sec, SSR output: 1.0 to 120.0 sec |
Manual reset | 0 to 100%( 0.0 to 100.0%) |
Relay life cycle_Mechanical | Min. 10,000,000 operations |
Relay life cycle_Electrical | Min. 100,000 operations(250VAC 3A resistance load) |
Memory Protection | Approx. 10 years(non-volatile semiconductor memory type) |
Insulation resistance | 100M??at 500VDC megger) |
Insulation type | Double insulation or reinforced insulation(mark: |
Dielectric strength | 1,000VAC 50/60Hz for 1 min(between input terminals and power terminals) |
Vibration | 0.75mm amplitude at frequency of 5 o 55Hz (for 1 min) in each X, Y, Z direction for 2 hours |
Noise immunity | ?0.5kV the square wave noise (pulse width: 1?? by the noise simulator |
Environment_Ambient temperature | -10 to 50?? storage: -20 to 60??/td> |
Environment_Ambient humidity | 35 to 85% RH, storage: 35 to 85% RH |
Protection structure | IP20(IEC standard) |
Accessories | Expansion connector: 1, module lock connector: 2 |
Approval | ![]() |
Weight_Basic module | Approx. 250.4g (approx. 177.3g) |
??hermocouple: Connecting 1 or more expansion module can vary measurement accuracy about ?1?? regardless of the number of connected expansion module.
??easured accuracy:
??t room temperature (23?????
??Thermocouple K, J, T, N, E below -100?? L, U, PLII, and RTD Cu50?, DPt50?: (PV ?0.3% or ?2?? higher one) ? 1-digit
??Thermocouple C, G and R, S below 200?? (PV ?0.3% or ?3?? higher one) ? 1-digit
??Thermocouple B below 400?? there is no accuracy standards.
??ut of room temperature range
??RTD Cu50?, DPt50?: (PV ?0.5% or ?3?? higher one) ? 1-digit
??Thermocouple R, S, B, C, G: (PV ?0.5% or ?5?? higher one) ? 1-digit
??Other blow -100?? within ?5??br>??he weight includes packaging. The weight in parenthesis is for unit only.
??nvironment resistance is rated at no freezing or condensation.
Tags: TMH4-N2CB